TECHNICAL SPECIFICATIONS Multiparameter Patient Monitor
- CE, ISO13485
- SFDA: Class ���b
- Anti-electroshock degree:Class��� equipment (internal power supply)
- TEMP/SpO 2 /NIBP: BF
- ECG/Resp: CF Quality Standards and Classification
- Adult/Pediatric/Neonatal���Necllor Spo2&Suntch Nibp���/Surgery/ Operating Room/ICU/CCU (Seperated parameter board) Application Range
- 12.1���TFT screen Display Operating environment :
- Temperature: 0 ~ 40 ���
- Humidity: 15% ~ 80%
- Pressure: 86.0 kPa ~ 106.0kPa Transport and Storage environment :
- Temperature: -20 ~ 55 ���
- Humidity: 10% ~ 90%
- Pressure: 86.0 kPa ~ 106.0kPa Environment
- AC: 100 ~ 240V, 50Hz/60Hz
- DC: Built-in rechargeable battery
- Battery: 11.1V 24wh Li-ion battery
- 2h operating after full charge
- 5min operating after low battery alarm Power Requirements
- Equipment : 305*162*290mm(about) ; 4kg(about)
- Package : 380*350*300mm(about) ; 6kg(about) Dimension and Weight
- Trend diagram/table: 720h
- NIBP review: 10000 events
- Wave review: 12h
- Alarm review: 200 alarm events
- Support drug concentration titration analysis Date Storage
- 5 Leads: RA, LA, LL, RL, V
- Lead mode: ���,���, ���, AVR, AVL, AVF, V ECG
- Gain: �� 1.25mm/mV,2.5mm/mV,
- 5.0mm/mV, 10mm/mV,
- 20mm/mV (error 5%)
- Sweep speed: 6.25mm/s,12.5mm/s,25mm/s,
- 50mm/s (error��� ��10%)
- Bandwidth: Surgery 1 ~ 25 Hz
- Monitor 0.5 ~ 40 Hz
- Diagnostic 0.05 ~ 130 Hz
- Sensitivity: >200 V P-P
- CMRR��� 112dB
- Differential Input Impedance: >5M
- ohm Electrode Offset Potential ��300mV
- Polarization Voltage �����500mV
- Input Impedance ���5 M
- Calibration Voltage: 1mV�� 5% Max. Mean Error 5mmHg Max. Standard Deviation: 8mmHg
- Overpressure protection: Adult 300mmHg Pediatric 240mmHg Neonatal 150mmHg
- ST Segment Range: -2.0mV ~ +2.0mV
- Accuracy: Ranging -0.8 mV ��� +0.8mV, measuring error is ��0.02mV or ��10%, The larger prevails. No definition for other ranges. ST Segment
- Method: impedance between RA-LL
- Resp Impedance range: 0.3 ~ 3
- Base Impedance range: 200 ~ 4000
- Bandwidth: 0.3 ~ 2.5Hz
- Gain: �� 0.25, �� 0.5, �� 1, �� 2, �� 4
- Resp rate:Measurement Range Adult 0 ~ 120BrPM Neonatal/Pediatric 0 ~ 150BrPM
- Resolution: 1BrPM
- Accuracy: 7 ~ 150BrPM ( 2BrPM) 0 ~ 6BrPM (unspecified)
- Apnea Alarm: 10 ~ 40s RESP
- Channel 2
- Measure & alarm range: 0 ~ 50 ���
- Resolution: 0.1 ���
- Accuracy (no sensor) ��0.1 ��� (0~50 ���) ���excluding sensor error���TEMP
- ECG cable & electrodes
- SpO2 sensor
- TEMP probe
- NIBP cuff & tube
- Power cord
- Lithium-ion battery
- User manual Standard accessories
- Etco2 (Optional)
- Dual IBP (Optional)
- Nellcor Spo2 (Optional)
- Suntch Nibp (Optional)
- Recorder (Optional) Optional
- Measurement range: 0 ~ 100%
- Resolution: 1%
- Precision: 70% ~ 100% ��2% 0%���69%���no definition given SpO2
- Measure/alarm range: 20 ~ 300bpm
- Resolution: 1bpm
- Accuracy ��3bpm Pulse Rate
- Method: Oscillometry
- Measure mode: Manual, Auto, STAT
- Measure interval in auto mode:1~5/10/15/30/60/90/120/ 180/240/480min
- Measure period in STAT mode: 5min
- Pulse rate range: 40 ~ 240bpm
- Measure & alarm range: Adult SYS 40 ~ 280mmHg DIA 10 ~ 220mmHg MEAN 20 ~ 240mmHg Pediatric SYS 40 ~ 220mmHg DIA 10 ~ 160mmHg MEAN 20 ~ 170mmHg Neonatal SYS 40 ~ 135mmHg DIA 10 ~ 100mmHg MEAN 20 ~ 110mmHg
- Static pressure measurement Range: 0 ~ 300mmHg Accuracy 3mmHg Resolution: 1mmHg Accuracy: